Applications

AnaLig Sr-01 particles may be used to concentrate Sr2+ from dilute samples prior to analysis or to separate Sr2+ from interfering substances. AnaLig? Sr-01 is pH independent.


Column Specifications

Mesh size 60-100
Binding capacity 0.1 - 0.3 mmol/g AnaLig?
Density 0.4 g/ml
Optimal Flow Rate 0.5 ml/g/minute
Feed Solution pH range < 0 - 9.5


SPE procedure

1. Wash column with Sr-free water or dilute acid.
2. Pass aqueous solutions through the column at specified flow rates.
3. Wash column with three bed volumes of Sr-free water or dilute acid.
4. Elute column in small volume of eluent (see elution section)
5. Wash column with Sr-free water and/or dilute acid to recycle.


Sample loading and washing volumes

The sample may be of any volume. However, not more than 10% of column capacity should be filled before eluting. Suggested minimum wash volumes are three bed volumes per gram of dry AnaLig (7.5 ml x dry AnaLig? weight). At least 5 ml are recommended for a one gram column.


Elution Step

AnaLig Sr-01 can be eluted with 0.03 M EDTA or NTA. Minimum elution volume is three bed volumes.


Column flow

For normal application, gravity flow should be adequate. Sample vessels may be attached to the column top. Slow flow rates are desirable; however, flow rates of 10 mL/min have been used with tightly packed columns. If flow is too slow, a Buchner flask and vacuum pump/aspirator may be applied. A stopcock may be used to adjust flow rate.


Competing matrix elements

The following elements will compete with Sr2+ for binding sites: Ba2+, Pb2+, K+ weakly.


Limitations and cautions

Because AnaLig Sr-01 contains organic complexing agents bound to silica beads, certain conditions should be avoided. Strong oxidizing agents such as aqua regia, H2O2 and Cl2 will destroy the complexing agents. F-, pH > 10, and high temperatures will destroy the silica beads.


Recycling and storage

AnaLig particles can be reused by following the Use Procedure above. Dry storage is recommended for extended periods of time. This is accomplished by vacuum drying the washed column, and placing the column in a cool, dry location with reduced light. Wet storage for shorter periods of time can be accomplished by filling the column with 0.01 M HNO3 or 0.01 M HCl.


Reference.

IBC Advanced Technology: AnaLig Data Sheet and Procedure Manual.




ウラニウム戦争 核開発を競った科学者たち
青土社
アミール・D・アクゼル

amazon.co.jpで買う
Amazonアソシエイト by ウラニウム戦争 核開発を競った科学者たち の詳しい情報を見る / ウェブリブログ商品ポータル


テーマ

注目テーマ 一覧


月別リンク

ブログ気持玉

クリックして気持ちを伝えよう!
ログインしてクリックすれば、自分のブログへのリンクが付きます。
→ログインへ

トラックバック(1件)

タイトル (本文) ブログ名/日時
「緊急時における食品の放射能測定マニュアル」詳細
厚生労働省が報道発表により紹介している緊急時の測定マニュアルを読み込んでおります。 詳細については、順次触れる予定ですが、今回は、取り急ぎ、放射性元素を分離するために利用される分離方が紹介されていたので、ポイントを抜粋して以下に記します。 ...続きを見る
必殺必中!前処理仕事人
2011/03/22 00:14

トラックバック用URL help


自分のブログにトラックバック記事作成(会員用) help

タイトル
本 文

コメント(0件)

内 容 ニックネーム/日時

コメントする help

ニックネーム
URL(任意)
本 文
Sr Separation by SPE resin with MRT 前処理仕事人/BIGLOBEウェブリブログ
文字サイズ:       閉じる